Application and Function
The Detachable Sampling Container SC 9R ensures safe sampling, especially in high temperature and high pressure applications such as steam boiler applications. It is also used to take samples in water, steam and condensation lines. The SC 9R sampling container is specially designed to meet all needs, including high pressure applications.
The detachable design makes it easy to clean, and the serpentines can be cleaned on the spot.
Conductivity measurement can be easily performed by taking a safe, accurate and precise boiler water sample with the SC 9R sample container.
Technical Data
Tip: SC 9R
Serpantin / Numune Tarafı
Nominal Pressure : PN 40
Bağlantı: 1/2” BSP – Dişli
Max. Çalışma Basıncı: 32 bar
Max. Çalışma Sıcaklığı : 239 °C
Gövde / Soğutma Tarafı
Nominal Pressure: PN 16
Bağlantı: 1/2” BSP – Dişli
Files
Catalog | Technical Data |
Max. Çalışma Basıncı: 16 bar
Max. Çalışma Sıcaklığı: 120 °C